Experimental Results for Colloidal Borescope Flowmeter
The colloidal borescope flowmeter (HB) used in this testing was developed by AquaVISION, Llc. The instrument employs a charge-coupled device camera, an optical magnification lens (140x), and a light-emitting diode (LED) illumination source to visually track neutrally buoyant colloids (approximately 1-5 microns) moving horizontally through the borehole. The device measured groundwater velocity and flow direction in the x-y (horizontal) plane. The HB magnetometer was checked with a compass to assure proper instrument orientation before deployment into the simulators. Borescope data were collected for each flow condition for a period of at least 30 minutes. After data collection, borescope data were inspected for anomalous periods of record that were caused by disturbances such as a vibration in the laboratory. Anomalous periods were deleted before computing the velocity and flow direction.
Complete Metadata
| bureauCode |
[ "010:12" ] |
|---|---|
| identifier | http://datainventory.doi.gov/id/dataset/USGS_5cc215ade4b09b8c0b74717b |
| spatial | -089.631599, 30.361283, -086.265765, 39.875856 |
| theme |
[ "geospatial" ] |